GE Medical Systems Information Technologies Inc recalls
2 recalls on record · FDA
Every recall we have on record for GE Medical Systems Information Technologies Inc, compiled from federal safety data. Check the hazard, the remedy, and how serious each one is, and always confirm on the official notice.
GE Healthcare MAC VU360, REF# part number 2030361-023; Electrocardiograph
Software issue issue with the MAC VU360 v3.0 system. After selecting the order refresh icon, an open preview or review screen might fail to close automatically after 2 minutes of inactivity. As a result, data from a previous patient may remain on the screen and could be incorrectly associated with a subsequent patient.
GE HealthCare TRADE IN CARESCAPE TELEMETRY SERVER ARK-2250L - NORTH AMERICA, REF…
GE HealthCare has become aware of an unlikely situation where potential loss of ECG and SpO2 monitoring involving ApexPro CARESCAPE Telemetry Server (CTS) v5.0 and v6.0, can result from a persistent NO COMM or OFF NETWORK condition at the CARESCAPE Central Station or CIC Pro Clinical Information Center (Central Station).