PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. 2026-01-21 Health serious FDA
Health · FDA · Recall #FDA-Z-1063-2026

Allura Xper FD10F; Model Number: 722002; Recall

Issued January 21, 2026 · PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.

In January 2026, PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalled the Allura Xper FD10F; Model Number: 722002;, a serious health recall reported by the FDA. Affected owners are eligible for a voluntary. Here is what the hazard is and exactly what to do about it.

Compiled by Maya Ellison, RecallCheckr

What this means for you

Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.

What you should do

  1. 1 Stop using Allura Xper FD10F; Model Number: 722002; right away.
  2. 2 Set it aside where it won't be used by mistake.
  3. 3 Contact PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. to arrange your voluntary. See the official notice below for contact details.
  4. 4 Keep the product until your remedy is fully processed. You may need it.
Official remedy (Voluntary):

Recall details

Brand
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Category
Health
Recall date
January 21, 2026
Units affected
Not reported
Sold at
Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Bangladesh, Belgium, Brazil, Cayman Islands, Chile, China,
Recall number
FDA-Z-1063-2026
Made in
Netherlands
Issuing agency
FDA